Efficient hybrid 3D system calibration for magnetic particle imaging systems using a dedicated device

Anselm von Gladiß, Matthias Gräser, André Behrends, Xin Chen & Thorsten M. Buzug
Image reconstruction in magnetic particle imaging is often performed using a system matrix based approach. The acquisition of a system matrix is a time-consuming calibration which may take several weeks and thus, is not feasible for a clinical device. Due to hardware characteristics of the receive chain, a system matrix may not even be used in similar devices but has to be acquired for each imager. In this work, a dedicated device is used for...