Application of the substitution procedure for the calibration of non-automatic weighing instruments : Expert Report DKD-E 7-1

The expert report is based on the calibration guideline DKD-R 7-2, Edition 01/2018 [German translation of the EURAMET Calibration Guide No.18 Version 4.0 (11/2015)]. The expert report contains additional requirements regarding the practical implementation of the substitution procedure, thus supporting the high competence required of an accredited calibration laboratory. The substitution procedure is described in chapter 4.3.3 of DKD-R 7-2. By using the substitution procedure, the measurement uncertainty contribution of the load used increases with...