Data from: Reliability of single-use PEEP-valves attached to self-inflating bags during manual ventilation of neonates – an in vitro study

Julia Christine Hartung, Silke Wilitzki, Marta Thio-Lluch, Arjan B. Te Pas, Gerd Schmalisch & Charles Christoph Roehr
Introduction International resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP. Methods Ten new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1kg neonatal manikin containing a lung model...