3 Works

EURADOS Intercomparison 2016 for Whole Body Dosemeters in Photon and Mixed Radiation Fields

Hannes Stadtmann, Andrew F. McWhan, Tom W.M. Grimbergen, Markus Figel, Ana M. Romero, B. J. Jansen, Christian Hranitzky & Christian Gärtner
Abstract: EURADOS Working Group 2 has developed a system for a self-sustained programme of regular intercomparisons (ICs), (Figel, M. Report to Council, WG02-SG2, 2007). The first four ICs, which were all for whole body dosemeters in photon fields, were carried out between 2008 and 2014 (IC2008, IC2010, IC2012, IC2014). In addition, three ICs with different scopes were carried out, two for extremity dosemeters in photon and beta fields (IC2009ext, IC2015ext) and one for whole-body dosemeters...

EURADOS Intercomparison 2019 for Extremity and Eye Lens Dosemeters - Summary of Procedures, Results and Conclusions

Andrew F. McWhan, Wioletta Dobrzynska, Hannes Stadtmann, Tom W. M. Grimbergen, Markus Figel, Ana M. Romero & Isabelle Clairand
Abstract: EURADOS Working Group 2 has developed a system for a self-sustained programme of regular dosemeter intercomparisons (ICs), (Figel, M. Report to Council, WG02-SG2, 2007). ICs for whole body dosemeters were carried out in 2008, 2010, 2012, 2014, 2016 and 2018. In addition, four ICs with different scopes were carried out, two for extremity dosemeters in photon and beta fields (IC2009ext and IC2015ext) and two for whole-body dosemeters in neutron fields (IC2012n and IC2017n). This...

Intercomparison IC2021area of passive area dosimetry systems

Christian Hranitzky, Christian Gärtner, Christian Naber, Julia Aslan, Maria A. Duch, Thomas Haninger & Željka Knežević
Abstract: The EURADOS intercomparison IC2021area was carried out between May 2021 and April 2022 for 66 participating passive H*(10) area dosimetry systems from 47 different institutes and monitoring services. Three measurement conditions were provided at locations of the Karlsruhe Institute of Technology: 3-months indoor, 3-months outdoor and 6-months outdoor. The challenge of this intercomparison was measuring additionally irradiated low dose radiation. Six dosemeters of each participating system were irradiated with Cs-137 gamma reference radiation: Three...

Registration Year

  • 2022
    1
  • 2021
    2

Resource Types

  • Data Paper
    3

Affiliations

  • Mirion Technologies (United States)
    3
  • Seibersdorf Laboratories (Austria)
    3
  • Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas
    2
  • Helmholtz Zentrum München
    1
  • Institut de Radioprotection et de Sûreté Nucléaire
    1
  • Rudjer Boskovic Institute
    1
  • Universitat Politècnica de Catalunya
    1
  • Karlsruhe Institute of Technology
    1